Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Soar 3100: (1 year) CAL VER /5500 DATE: 10-Sep-97 AUTHOR: Fluke Corporation REVISION: $Revision: 1.5 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 17 NUMBER OF LINES: 90 CONFIGURATION: Fluke 5500A ============================================================================= # # Compatibility: # 5500/CAL 4.0 or MET/CAL 4.0 or later. # # Subprocedures: # None # # Required Files: # 3100.bmp # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5500A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 DISP {Procedure not formally tested with UUT.} 1.001 DISP {See procedure README file.} 1.002 ASK- P F V 1.003 ASK+ K X 1.004 HEAD PRELIMINARY INSTRUCTIONS 1.005 DISP Warm-up time: NONE 1.005 DISP Temperature: 0 - 40 degC 1.005 DISP Humidity: <70% 1.006 DISP [32] WARNING 1.006 DISP HIGH VOLTAGE is used or exposed during the performance 1.006 DISP of this calibration. DEATH ON CONTACT may result if 1.006 DISP personnel fail to observe safety precautions. 1.007 PIC 3100 1.008 DISP Set the POWER switch to ON. 1.009 HEAD {DC VOLTAGE} 1.010 HEAD DC VOLTAGE: {200mV Range} 1.011 5500 200 100.0mV 0.5% 0.3U 2W 2.001 HEAD DC VOLTAGE: {2000mV Range} 2.002 5500 2000 1000mV 0.7% 2U 2W 3.001 HEAD DC VOLTAGE: {20V Range} 3.002 5500 20 10.00V 0.7% 0.02U 2W 4.001 HEAD DC VOLTAGE: {200V Range} 4.002 5500 200 100.0V 0.7% 0.2U 2W 5.001 HEAD DC VOLTAGE: {500V Range} 5.002 5500 500 500V 0.7% 2U 2W 6.001 HEAD {AC VOLTAGE} 6.002 HEAD AC VOLTAGE: {2000mV Range} 6.003 5500 2000 1000mV 1% 5U 400H SI 2W 7.001 HEAD AC VOLTAGE: {20V Range} 7.002 5500 20 10.00V 1% 0.05U 400H SI 2W 8.001 HEAD AC VOLTAGE: {200V Range} 8.002 5500 200 100.0V 1% 0.5U 60H SI 2W 9.001 HEAD AC VOLTAGE: {500V Range} 9.002 5500 500 500V 1% 5U 60H SI 2W 10.001 HEAD {RESISTANCE} 10.002 HEAD RESISTANCE: {20 MOhm Range} 10.003 5500 20 10.00MZ 2% 0.03U 2W 11.001 HEAD RESISTANCE: {2000 kOhm Range} 11.002 5500 2000 1000kZ 1.2% 2U 2W 12.001 HEAD RESISTANCE: {200 kOhm Range} 12.002 DISP Connect the 5500A to the UUT as follows: 12.002 DISP [32] 5500A AUX HI to UUT Probe Tip 12.002 DISP [32] 5500A AUX LO to UUT COM 12.002 DISP [32] 5500A NORMAL HI to UUT Probe Tip 12.002 DISP [32] 5500A NORMAL LO to UUT COM 12.002 DISP Connect the SENSE leads DIRECTLY to the UUT. Connect 12.002 DISP the NORMAL leads to the SENSE leads AT the UUT. 12.003 DISP Set the FUNCTION switch to OHMS. 12.004 5500 200 100.0kZ 0.7% 0.2U CW 13.001 HEAD RESISTANCE: {20 kOhm Range} 13.002 5500 20 10.00kZ 0.7% 0.02U CW 14.001 HEAD RESISTANCE: {2000 Ohm Range} 14.002 5500 2000 1000Z 0.7% 2U CW 15.001 HEAD RESISTANCE: {200 Ohm Range} 15.002 5500 200 100.0Z 0.7% 0.4U CW 16.001 HEAD {CONTINUITY} 16.002 DISP Press the OHMS/TONE selector located on the top 16.002 DISP of the unit, and check the display for "*))". 16.003 5500 100Z S CW 16.004 EVAL Is {there an AUDIBLE TONE present}? 17.001 END